26. Equipment and materials

Version 2.0

Refer to principles 7 and 8

Regulatory principles that apply to licensed centres

  • 7.

    conduct all licensed activities with proper skill and care and in an appropriate environment, in accordance with good clinical practice, to ensure optimum outcomes and minimum risk for patients, donors and offspring;

  • 8.

    ensure that all premises, equipment, processes and procedures used in the conduct of licensed activities are safe, secure and suitable for purpose;

…see all the principles

Close

Expand

Search the Code

Search term not found

Human Fertilisation and Embryology (HFE) Act 1990 (as amended)

17 Person responsible

  • (1)

    It shall be the duty of the individual under whose supervision the activities authorised by a licence are carried on (referred to in this Act as the "person responsible") to secure—



    (b) that proper equipment is used,

Licence conditions

  • T22
    For every critical activity, identifying information about all of the materials and equipment must be documented.
  • T23
    Activities must be carried out using equipment and materials designated for the purpose and maintained to suit their intended purpose and must minumise any hazard to patients and/or staff.
  • T24
    All critical equipment and technical devices must be identified and validated, regularly inspected and maintained in accordance with the manufacturer’s instructions. Where equipment or materials affect critical processing or storage parameters (eg, temperature, pressure, particle counts, microbial contamination levels) they must be identified and be the subject of appropriate monitoring, alerts, alarms and corrective action, as required, to detect malfunctions and defects, and to ensure that the critical parameters are maintained within acceptable limits at all times. All equipment with critical measuring function must be calibrated against a traceable standard if available.
  • T25
    New, repaired and recommissioned equipment must be tested and validated before use. Test results must be documented.
  • T26
    Maintenance, servicing, cleaning, disinfection and sanitation of all critical equipment and premises must be performed regularly and recorded accordingly.
  • T27
    Procedures for the operation of each piece of critical equipment must be established and these procedures must document the action to be taken in the event of malfunctions or failure.
  • T28
    Sterile instruments and devices must be used for the procurement of gametes and embryos. Instruments or devices must be of good quality, validated or specifically certified and regularly maintained for the procurement of tissues and cells.
  • T29
    When reusable instruments are used, a validated cleaning and sterilisation procedure for removal of infectious agents has to be in place.
  • T30
    Wherever possible only CE marked medical devices must be used.
  • T31
    The procedures for licensable activities must detail the specifications for all critical materials and reagents. In particular, specifications for additives (eg, solutions) and packaging materials must be defined. Critical reagents and materials must meet documented requirements and specifications and, when applicable, the requirements of Council Directive 93/42/EEC of 14 June 1993 concerning medical devices and Directive 98/79/EC of the European Parliament under the Council of 27 October 1998 on In vitro Diagnostic Medical Devices.

Back to top

Scope

  • 26.1
    For the purpose of this Code of Practice, 'equipment and materials' includes all equipment, disposables, reagents, and calibrations and control materials used in the conduct of assisted conception process.

Protection and hygiene

  • 26.2
    The centre should provide proper clothing and equipment for the personal protection and hygiene of staff carrying out licensed activities, together with written instructions for their use.

Managing equipment and material

  • 26.3

    The centre should establish documented procedures for managing equipment and materials, including:

    a) selecting and procuring equipment and materials

    b) ensuring the traceability of any products or materials that come into contact with gametes or embryos and that affect their quality and safety, and

    c) maintaining inventory information and records for stock control.

    d) ensuring software-driven equipment is effectively validated, and revalidated
    after any software update.

CE marking

  • 26.4

    The centre should use only media and consumables that have been CE-marked at a classification suitable for their intended purpose. Modifying existing devices (for example, adding calcium ionophore to culture medium) or using them ‘off label’ for purposes not intended by the manufacturer (for example, using a medium for a different purpose from that specified) has safety implications. It may also count as manufacture of a new device under the Medical Devices Regulations.

  • 26.5
    If the centre does choose to modify an existing product or use a product ‘off label’, it should (as the ‘manufacturer’) complete a risk analysis and validation to ensure the product or process is safe.

Safety of equipment used to store cryopreserved gametes and embryos

  • 26.6

    All centres storing gametes and embryos should have effective alarms and monitoring systems to ensure the safety of cryopreserved gametes and embryos. These systems should have:

    a) local alarms (ie, on individual dewars for either temperature or liquid nitrogen level)

    b) an auto-dial facility or similar (eg, link to fire-alarm board) to contact staff outside normal working hours

    c) adequate staffing and funding to implement formal emergency procedures, including having on-call arrangements, and

    d) adequate spare storage space or vessels to enable transfer of samples if a vessel fails.

Back to top